Flow Meter Calibration
The flow meter supplied with your system has been calibrated at the factory and
under normal circumstances there may be no need to re-calibrate it. However,
the factory calibration setup may not reflect specific sprayer plumbing. Before
spraying actual chemicals, the flow meter should be checked for proper
calibration. Also, the flow meter is made up of moving parts and can wear over
time. If you feel that your flow meter is worn, you can re-calibrate your flow meter
using the 844 console. To do so, use either of the following methods:
Method 1 – Known Volume
Step 1 – Known Value
Select a known volume of water (n) that you will pump through the flow meter. In
these steps, 10 US gallons (38 liters) {8 Imperial gallons} will be used as an
example. The specific volume you use is not important, although larger volumes
generally produce more accurate results.
Step 2 Programming Calibration Number
Go into the System Setup mode and advance to the flow meter calibration number
step (flashing turbine at the top of the display). In this programming step, enter (n)
10 US gallons (38 liters) {8 Imperial gallons} as the calibration number. Exit the
System Setup mode by depressing and holding the key for five seconds.
Step 3 Resetting Volume Counter
Reset the total volume number that is displayed in the lower right portion of the
display. (Display alternates between Ac (Ha) covered and the volume sprayed through
the flow meter). To reset this number, press and hold the and keys
simultaneously for three seconds or until you see the values reset to zero.
Step 4 Spraying Known Volume
Engage the pump on the sprayer, flip the Master boom section of the 844 to ON
(make sure at least one boom section switch is on), and pump exactly (n) 10 US
gallons (38 liters) {8 Imperial gallons} through the flow meter. When the known
volume is reached, shut the Master boom switch OFF.
Step 5 Entering in New Flow meter Calibration Number
The number that is displayed in the lower right corner of the display is your new
flow meter calibration number. Record this number, enter the System Setup
mode and advance to the flow meter calibration number step. Enter the new
value from your test into this step and then exit the System Setup mode.
Step 6 Double-checking the New Value
Double-check your new value by spraying a larger volume of water, such as
100 US gallons (378 liters) {83 Imperial gallons} through the flow meter. Before
spraying the larger volume, make sure to reset the memory counters by
depressing and holding the and keys simultaneously for three seconds.
NOTE: After double checking the new calibration number, if you find that
the volume shown on the display of the 844 is more than +/- 3% in error,
change the flow meter calibration number in the System Setup mode by
the % difference it is in error.